Description
■ Caution! This device is for industrial use, not for clinical use.
■ What is the distance-to-spot ratio?
A distance-to-spot ratio, which is the ratio of the Distance between the object and the thermometer to the size of the Spot, is preset at D:S = 8:1 for SK-8700. For example, if the distance is 320 mm, a spot circle of diameter 40 mm can be measured. Make sure that the object is larger than the circle.
* The spot circle is defined as an area to receive 90% of the emitted energy from the object.
■ Warning: Do not look into the laser marker source or point it on eyes as it may damage eyes.
SK-8700 conforms to the Class II in the Standard for Safe Use of Laser Products (JIS C6802) which is defined as follows: A visible light (wavelength between 400 and 700 nm) with an output level less than 1 mW that a human’s protective response can resist.
PSC mark is labeled on the device to prove its quality based on Japan’s Consumer Product Safety Act.
Feature
■ Enables clean contactless measurement which is ideal for food industries where hygienic measurement is required.
■ One of the best advantages of a contactless infrared thermometer is its quick response. With SK-8700 you can easily and instantly measure surface temperature in one hand.
■ Applicable for a wide range of temperature from −20 to 420°C (0 to 786°F). Using a single-point laser marker helps you identify the target at a glance.
■ Certified for PSC, a standard based on Japan’s Consumer Product Safety Act, as a class II portable laser device.
■ Turns off the power automatically if the keys are not pressed for 5 seconds, to prevent from running out the battery even when you forget to turn it off.
■ Auto hold function keeps the measured value on the LCD for 5 seconds.
■ Equipped with a lens cap to protect the sensing section from dust or foreign matter. It also protects users from looking at laser activated accidentally when not used.
■ Traceable calibration available
■ °C/°F displays interchangeable
Specifications
Cat. No. | 8260-00 |
Product name | Infrared Thermometer with Laser Marker |
Model name | SK-8700 |
Measuring range | −20 to 420°C (0 to 786°F) |
Measurement accuracy | ±2°C (−20 to 100°C / 0 to 212°F) ±2%rdg (100 to 420°C / 212 to 786°F) Under blackbody calibration with ambient temperature 23±5°C / 73±9°F, emissivity 0.95 |
Resolution | 1°C / 1°F |
Operation ambient | 0 to 50°C / 32 to 112°F, less than 80%rh without condensing |
Detector | Thermopile |
Emissivity | Fixed at 0.95 |
Distance coefficient | D:S = 8:1 (where D is distance, S is diameter of spot circle) |
Repeatability | ±1°C |
Measured wavelength | 6 to 14 μm |
Response time | 0.5 sec. |
Laser marker | Single-point, 650 nm, less than 1 mW |
Power consumption | 40 mA max. when laser marker and backlight are turned on |
Power | 9-volt battery x1 |
Battery life | Approx. 5 hours if both laser and backlight are on 10 hours if either laser or backlight is on 50 hours if both laser and backlight are off |
Material | ABS resin |
Dimensions | 41.5 (W) x 160 (H) x 65 (D) mm |
Weight | 140 g including a battery |
Standard accessories | Hand strap with lens cap x1, vinyl cover x1, vinyl pouch x1, 9-volt zinc-carbon battery x1, instruction manual |